Neonatal Incubator Infant Radiant Warmer
Why is a newborn placed under a radiant heat warmer?
Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized.
Who needs radiant warmer?
Any infant who weighs less than 1000 grams should be cared for on a radiant warmer or in a servocontrolled double-walled incubator.
Features of Infant Radiant Warmer MERW-4000A/B:
1. LCD TOUCH SCREEN Temp. servo-controlled by computer;
2. Separate setting temp., real temp. and timer display screens;
3.Infrared quartz heater
4.Available adjust of radiator and mattress tilting
5.X-ray tray under the bed
6.Four self-check alarms
7.RS-232 connector APGAR timer
Note: Based on MERW-4000A, added with suction,Oxygen bottle (6.3L*2) and Oxygen flow meter
Specifications Infant Incubator Warmer:
Model | MERW-4000A | MERW-4000B |
Electrical requirement | ~220V50Hz or~230V60Hz |
Power consumption | ≤800VA |
Air mode temp. range | 32℃-37.5℃ |
Over temp. alarm | <38.5℃ |
Temp. deviation alarm | ±1.0℃ |
Skin sensor accuracy | ≤±0.3℃ |
Mattress temp. uniformity | ≤2.0℃ |
Heating rate | ≤30min |
APGAR timing | 1min,5min,10min |
Model | MERW-4000A | MERW-4000B |
Screen | LCD |
Skin temperature senor | ● | ● |
Suction | - | ● |
O2 bottle | option |
Electronic lift | option |
Phototherapy | option |